background image

 

 

 

Fitting Deuterium Lamp Background Correction to 
Agilent 50/55 AA Instruments 

 
This instruction sheet describes how to install a deuterium lamp background 
correction upgrade to an Agilent 50/55 AA. 

Parts List 

Description Part 

Number 

Description Part 

Number 

Assembly, mirror M5D2 

0110598890 

Screw (2 off) M5x30 countersunk 

1510187300 

Assembly, beamsplitter 

0110599390 

Screw (3 off) M4x12 pan 

1510108200 

Assembly, D

2

 lamp 

0110602800 

Set screw (2 off) M4x16 cone point  1510180800 

Assembly, loom D

2

 lamp 

0110613000 

Set screw (2 off) M4x16 oval point 

1510180900 

Ball bearing 

0310031100 

Washer 

1510126900 

Bracket 

0410293800 

Hex nut M5 

1510110200 

Bracket, microswitch cover 

0410309700 

Window, UV silica 

2010082600 

Microswitch mounting plate 

0810018700 

Label, D

2

 interlock 

3110050300 

Cover, D

2

 lamp 

0910106900 

Spring (2 off) 

4610016600 

Mount plate 

0910110700 

Adhesive pad 

4910012700 

Screw (2 off) 8 BA 

1510062600 

Installation instructions 

8510160900 

Screw M4x30 

1510109300 

Insulating washer 

21100002 

Installation 

To fit the deuterium lamp background correction upgrade:  

1

 

Unpack the contents of the kit (part number 9910092500). 

2

 

Unplug the 50/55 AA instrument. 

 

WARNING 

 

 

Electrical Shock Hazard 
The instrument contains electrical circuits, devices and components that operate at dangerous 
voltages. Contact with these circuits, devices or components can cause death, serious injury or 
painful electric shock. 

 

3

 

Remove external covers and the optics cover as described in the Agilent 
50/55/140/240/280 AA Service Manual. 

 

 