Fitting Deuterium Lamp Background Correction to
Agilent 50/55 AA Instruments
This instruction sheet describes how to install a deuterium lamp background
correction upgrade to an Agilent 50/55 AA.
Parts List
Description Part
Number
Description Part
Number
Assembly, mirror M5D2
0110598890
Screw (2 off) M5x30 countersunk
1510187300
Assembly, beamsplitter
0110599390
Screw (3 off) M4x12 pan
1510108200
Assembly, D
2
lamp
0110602800
Set screw (2 off) M4x16 cone point 1510180800
Assembly, loom D
2
lamp
0110613000
Set screw (2 off) M4x16 oval point
1510180900
Ball bearing
0310031100
Washer
1510126900
Bracket
0410293800
Hex nut M5
1510110200
Bracket, microswitch cover
0410309700
Window, UV silica
2010082600
Microswitch mounting plate
0810018700
Label, D
2
interlock
3110050300
Cover, D
2
lamp
0910106900
Spring (2 off)
4610016600
Mount plate
0910110700
Adhesive pad
4910012700
Screw (2 off) 8 BA
1510062600
Installation instructions
8510160900
Screw M4x30
1510109300
Insulating washer
21100002
Installation
To fit the deuterium lamp background correction upgrade:
1
Unpack the contents of the kit (part number 9910092500).
2
Unplug the 50/55 AA instrument.
WARNING
Electrical Shock Hazard
The instrument contains electrical circuits, devices and components that operate at dangerous
voltages. Contact with these circuits, devices or components can cause death, serious injury or
painful electric shock.
3
Remove external covers and the optics cover as described in the Agilent
50/55/140/240/280 AA Service Manual.