Chapter 3: Testing Performance
Offset Performance Test
3–5
Offset Performance Test
Specifications
Equipment Required
Procedure
Zero Error Test
1
Disconnect all cables from the scope channel inputs.
2
Press Default Setup, then configure the scope as follows:
a
Pull down the Setup menu and select Acquisition.
b
When the Acquisition Setup window is displayed, enable averaging and set the # of
averages to 256 as shown below.
Offset Accuracy
= ±(2% of channel 1% of full scale)
= ±(Offset Gain + Zero Error)
Offset Range
> ±12 div, or ±4 V, whichever is smallest
Full scale is defined as 8 vertical divisions. Vertical divisions are defined by the major scale settings above non-major scale settings.
The major scale settings are 10 mV, 20 mV, 50 mV, 100 mV, 200 mV, 500 mV and 1 V.
Description
Critical Specifications
Recommended Model/
Part Numbers
Digital Multimeter
DC voltage measurement accuracy better than ±0.1% of reading Agilent 34401A or
Agilent 3458A
Cable Assembly
(2 required)
50
Ω
characteristic impedance, BNC (m) connectors
Agilent 8120-1840
Adapter
BNC Tee (m)(f)(f)
Agilent 1250-0781
Adapter
BNC (f) to dual banana
Agilent 1251-2277
Summary of Contents for 54853A
Page 4: ...iv Rear panel with option 017 Removable hard drive ...
Page 6: ...vi ...
Page 10: ...Contents 4 ...
Page 24: ...1 14 ...
Page 44: ...2 20 ...
Page 172: ...6 38 ...
Page 177: ...Chapter 7 Replaceable Parts Exploded Views 7 5 Exploded Views Front Frame and Front Panel ...
Page 178: ...Chapter 7 Replaceable Parts Exploded Views 7 6 Fan and Acquisition Assembly ...
Page 181: ...Chapter 7 Replaceable Parts Exploded Views 7 9 Sleeve and Accessory Pouch ...
Page 187: ...8 Block Level Theory 8 3 Acquisition Theory 8 5 Theory of Operation ...
Page 192: ...8 6 ...