15
Chapter 1: Operating the Probe
Probe Operating Range
Using the probe with the 1143A power module
Probe power and offset control is provided by the 1143A Probe Offset Control
and Power Module.
1
Set up the power module by following the instructions in the User and
Service Guide.
2
Connect the probe using “To connect the probe” on page 12" of this
guide.
3
Turn on the power for the power module.
4
Set the appropriate Remote/Local switch.
•
To control offset voltage with the power module, set the switch to
Local.
•
To control the offset voltage remotely, set the switch to Remote.
5
With Local control, set the appropriate Zero/Variable switch.
•
To enable the local offset control, set the switch to Variable.
•
To disable the local offset control, set the switch to Zero.
6
Connect the probe to the signal to be measured.
If the oscilloscope has an offset feature, be sure that it is set to zero so
that the probe offset does not have to compensate for the oscilloscope
offset.
7
If necessary, adjust the Coarse and Fine offset controls so the desired
part of the signal is displayed on the oscilloscope. See sidebar below.
The offset range is greater than 50 V relative to the probe tip.
Bandwidth issues are the same as covered on the previous page
See Also
The
User and Service Guide
for the 1143A Probe Offset Control and Power
Module about remote probe operation.
The probe has limiting designed to avoid excessive power dissipation. The input
operating range of the probe is 5 V. If the input and offset e14V relative
to the probe tip, the output of the probe will limit at +1.4 V. As the input plus
offset reaches -14 V the output will limit at -1.4 V; then, it will fold back to
approximately -0.8V as the input plus offset exceeds -14 V. The output of the
probe will remain at the limit voltage until the input plus offset falls below
approximately -8 Vdc.