Triggering the Oscilloscope
To use LIN triggering
To use LIN triggering
LIN (Local Interconnect Network) trigger setup consists of connecting the
oscilloscope to a serial LIN signal.
The LIN trigger will trigger on the rising edge at the Sync Break exit of the LIN
single-wire bus signal that marks the beginning of the message frame. A LIN
signal message frame is shown below:
1 Press the
More
key in the Trigger section of the front panel, rotate the
Entry knob until
LIN
is displayed in the
Trigger
softkey, then press the
Settings
softkey to display LIN trigger menu.
2 Press the
Baud
softkey to set the LIN signal baud rate to match your LIN
bus signal.
The LIN baud rate can be set to 2400 b/s, 9600 b/s, or 19.2 kb/s. The default
baud rate is 2400 b/s.
Sync Break Exit
Sync
Break
Sync
Field
Identifier
Break
Data
Fields
Checksum
Field
Signal
baud rate
Trigger on:
condition
Signal type
Signal
source
Trigger level
or threshold
Currently selected
signal source
LIN trigger
Return to
previous menu
Summary of Contents for 54621A
Page 4: ...iv ...
Page 10: ...Contents 6 ...
Page 11: ...1 Getting Started ...
Page 38: ...1 28 ...
Page 39: ...2 Front Panel Overview ...
Page 63: ...3 Triggering the Oscilloscope ...
Page 117: ...4 MegaZoom Concepts and Oscilloscope Operation ...
Page 171: ...5 Making Measurements ...
Page 222: ...5 52 Making Measurements Making overshoot and preshoot measurements ...
Page 223: ...6 Utilities ...
Page 239: ...7 Performance Characteristics ...
Page 250: ...7 12 Performance Characteristics Agilent 54620 series Performance Characteristics ...
Page 260: ...7 22 ...