Technical Information
Start-Up
5
Agilent 4UHV Ion Pump Controller User Manual
/
87-900-137-01 (A)
67/104
Start-Up
This chapter explains how to switch on and off the unit and the high
voltage channels. In the following chapters, all the functions of the
controller are explained in detail.
1
Put the main power switch situated on the front panel to OFF
2
Insert the power cable into the correct connector on the rear
panel
3
Connect the HV cables to the controller and to the ion pumps
4
Make sure that “Interlock” of HV cable is connected
5
To switch on the unit, put the main power switch to ON
6
Select Local for the Operating mode
7
Press the keys HV-ON/OFF keys and HV-x, where x is the number
of the channel you want to switch on.
Figure 12
Front Panel
Main power switch
Keys HV-x
Key HV-ON/OFF