Appendix B
Information for Replacing 4278A with 4288A
Functional Comparison and GPIB Command Correspondence
Integration time (measurement time)/averaging
Functional comparison
Differences in usage of GPIB commands
GPIB command correspondence
Cable length
Functional comparison
GPIB command correspondence
4278A
4288A
Integration time (measurement time)
SHORT, MEDIUM, LONG
SHORT
*1
, LONG
Averaging count
Allowable setup range
1 to 256
1 to 256
Resolution
Only 9 points:
1,2,4,8,16,32,64,128,256
1
*1.The SHORT mode of the 4288A corresponds to both the SHORT and MEDIUM modes
of the 4278A.
4278A
4288A
ON/OFF of the
averaging function
The averaging function cannot be turned
ON/OFF. It is always ON (equivalent to
OFF when the averaging count is set to 1).
The averaging function can be turned
ON/OFF. You have to turn ON the
averaging function when setting the
averaging count.
Table B-5
Correspondence of measurement time/averaging/trigger delay time setup
commands
Setup
GPIB command
4278A
4288A
Integration time
(measurement time)
SHORT
ITIM1
:APER SHOR
MEDIUM
ITIM2
LONG
ITIM3
:APER LONG
Averaging count
AVE=
4278A
4288A
Cable length
0 m, 1 m, 2 m
0 m, 1 m, 2 m
Table B-6
Correspondence of setup commands related to measurement time
Setup
GPIB command
4278A
4288A
Cable length
0 m
CABL0
1 m
CABL1
2 m
CABL2
Summary of Contents for 4288A
Page 18: ...16 Chapter1 Making Effective Use of This Manual How To Use This Manual ...
Page 212: ...210 AppendixA Manual Changes Manual Changes ...
Page 248: ...246 AppendixB Information for Replacing 4278A with 4288A Comparison of Interfaces ...
Page 258: ...256 AppendixC Status Reporting System Using the Status Reporting System ...
Page 288: ...286 Index Index ...