4-16
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Making a Measurement
Sweep Measurements
To Set up Secondary Sweep
Secondary sweep source is the measurement unit defined as VAR2 in the
CHANNELS: CHANNEL DEFINITION screen. On the MEASURE: SWEEP
SETUP screen, set up the primary sweep source (VAR1), then do following:
1. VAR2: START
Enter the secondary sweep start value.
2. VAR2: STEP
Enter the secondary sweep step value.
3. VAR2: NO OF STEP
Enter the number of steps for the secondary sweep.
4. VAR2 : COMPLIANCE, POWER COMPLIANCE
Only for SMU. Enter the compliance value, and power compliance (ON or OFF)
for the secondary sweep source.
SWEEP MODE and LIN/LOG are automatically set to SINGLE and LINEAR,
respectively. You cannot change the settings.
The STOP value is automatically calculated from START, STEP, and NO OF STEP
values.
You cannot change UNIT and NAME in this screen. To change the values, go to the
CHANNELS: CHANNEL DEFINITION screen.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...