Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
3-67
Measurement Functions
Zero Offset Cancel
Table 3-10
Ranging Mode Available for Offset Measurement
Table 3-11
Allowable Offset Value
Measurement Mode
Unit
Available Ranging
Mode
Measurement
Range
a
a. Offset data is measured in the measurement range shown above.
Current Measurement
HPSMU,
MPSMU
auto
1 nA
1 nA limited or fixed
1 nA
HRSMU
auto
10 pA
10 pA limited or fixed
10 pA
100 pA limited or fixed
100 pA
1 nA limited or fixed
1 nA
Differential Voltage
Measurement
VMU
auto, limited auto, fixed
0.2 V
b
b. VMU2 measures voltage in grounded mode to confirm that voltage does
not exceed ±20 V.
Measurement
Mode
Unit
Measurement
Range
Allowable Offset Value
Current
Measurement
HPSMU
1 nA
less than ±1 % of range
MPSMU
1 nA
less than ±1 % of range
HRSMU
1 nA
less than ±1 % of range
100 pA
less than ±1 % of range
10 pA
less than ±4 % of range
Differential Voltage
Measurement
VMU
0.2 V
less than ±1 % of range
a
a. If VMU2 grounded mode measurement value is greater than or equal to
±20 V, error occurs.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...