3-54
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
Measurement Ranging Mode
Auto Ranging
The monitor unit automatically searches for and measures at the range that provides
the highest resolution as follows:
V measurement
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 110 % of the range, then the unit performs the
measurement.
I measurement
•
1 A to 1
m
A
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
If the measurement value is less than 1 % of the present range and the present
range is 100
m
A or higher range, the range changes down two ranges instead of
one range.
•
100 nA to 100 pA
The unit changes ranges (up or down one range at a time) until the measurement
value is between 10 % and 114 % of the range, then the unit performs the
measurement.
•
10 pA
The unit changes to the next higher range when the measurement value exceeds
104 % of the present range.
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...