3-48
Agilent 4155B/4156B User’s Guide Vol.2, Edition 5
Measurement Functions
R-BOX Control
R-BOX Control
Agilent 16441A R-Box must be used for applications which need to connect a series
resistor between SMU and DUT. For example, the R-Box is effective for the DUT
protection. If sudden voltage change occurs at DUT, excessive current flows to the
DUT, and it may damage the DUT without the R-Box. In other case, you may want
to measure negative resistance characteristics. This application needs series resistor
because SMUs cannot measure negative resistance.
The 4155B/4156B automatically compensates for voltage drop of the series
resistance value. So, the GRAPH/LIST screen group show the compensated data.
For details of the 16441A R-Box, refer to
Agilent 16441A R-Box User's Guide
.
Resistance Value
The 16441A R-Box provides two sets of the series resistors. So you can connect two
SMU channels to the R-Box. Resistance values of resistors are listed below:
•
1 M
W
•
100 k
W
•
10 k
W
•
0
W
You can select the resistance values on the 4155B/4156B setup screen. The LEDs on
the 16441A R-Box indicate the present resistance value.
Connections
Following table is the parts list of cables for connecting the 16441A R-Box.
Agilent Model or Part Number
Description
04155-61610
Control Cable 1.5 m
04155-61609
Control Cable 3.0 m
04155-61605
Triaxial Cable 0.4 m
16493K #001
Kelvin Triaxial Cable 1.5 m
16493K #002
Kelvin Triaxial Cable 3.0 m
Summary of Contents for 4156b
Page 18: ...Contents 10 Agilent 4155B 4156B User s Guide Vol 2 Edition 5 Contents...
Page 19: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 1 Measurement Units...
Page 41: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 2 Measurement Mode...
Page 75: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 3 Measurement Functions...
Page 143: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 4 Making a Measurement...
Page 205: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 5 Analyzing Measurement Results...
Page 243: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 6 Screen Organization...
Page 333: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 7 Data Variable and Analysis Function...
Page 375: ...Agilent 4155B 4156B User s Guide Vol 2 Edition 5 8 If You Have A Problem...