Startup
5
User Manual
39
Verify 35900E Series II is Ready for Operation
At power up, the 35900E Series II automatically executes internal self-tests to
determine that everything is operating properly. These tests take about one
second.
When the 35900E Series II has completed the internal self-tests, verify that:
•
The
FAULT
LED is OFF.
•
The
POWER
LED is ON.
•
No other indicators are blinking.
If the red
FAULT
LED is blinking about once per second, the 35900 Series II is
configured for DHCP and the external computer has not provided an IP
address. Check the external networking procedures (See
You can use the test chromatogram stored in the 35900E Series II ROM in
conjunction with the host computer software to check the operation of the
entire system. The test chromatogram is turned on via configuration switches
3 through 6 on the 35900E Series II's back panel.