Trigger Delay
You can insert a delay between the trigger signal and each sample that
follows. This may be useful in applications where you want to allow the
input to settle before taking a reading, or for pacing a burst of readings.
If you do not specify a trigger delay, the meter automatically selects a
delay for you.
•
Delay range: 0 to 3600 seconds.
The default trigger delay is
automatic; the delay is determined by function, range, and integration
time (see also “Automatic Trigger Delays,” on page 89).
•
The trigger delay is stored in
volatile
memory; the meter selects an
automatic trigger delay when power has been off or after a remote
interface reset. The
CONFigure
and
MEASure?
commands set the
trigger delay to automatic.
•
If you specify a delay other than automatic,
that same delay
is used for
all
functions and ranges.
•
If you have configured the meter to take more than one reading per
trigger (sample count > 1), the specified trigger delay is inserted
between the trigger and
each
reading.
•
Front Panel operation:
You can use an automatic trigger delay or
you can specify a delay in seconds. Set the delay time in the
2: TRIG DELAY
command of the TRIGger menu.
If an automatic trigger delay is enabled, “AUTO” is displayed
momentarily before the actual number of seconds is displayed.
AUTO
3
Chapter 3 Features and Functions
Triggering
87
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......