Voltage Measurement Configuration
Input Channels
The meter has two independent input channels for measuring dc Volts.
You can make measurements on either or both channels, measure the
difference between the channels, or measure the ratio between the
channels. Each channel has an independent settings for null and range.
The figure on page 19 shows connections for two input channels.
•
Channel 1
is the default input channel. Both channels use the same
integration time setting.
•
The maximum measurable input on
Channel 1
is 120V, on
Channel 2
is 12 V.
•
For two channel operation, the maximum voltage difference between
Channel 1
LO and
Channel 2
LO is
±
150 V.
•
For either channel, the maximum non-destructive input is
±
150 V.
•
If you enable input filters, the filters are active for both input
channels. Switching between the channels will reset the digital filter
reading count (
see page
Channel
•
Front Panel operation:
Toggle the input channel by pressing the
Channel key on the front panel. The
CH1
or
CH2
annunciator will
light in the display. The meter keeps range and null settings
independently for each channel.
Scan
Shift
Channel
Choose the scan mode to alternately show the
CH1
and
CH2
readings.
Disable the digital filter when using the scan mode. In scan mode, the
meter alternates between the
CH1
and
CH2
readings every 3 seconds
or for each measurement (whichever is longer).
Chapter 3 Features and Functions
Voltage Measurement Configuration
66
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......