Input Filter Commands
See also page 55 in Chapter 3.
Using the input filters from the remote interface
is not
recommended. To
use the filters with the remote interface be aware of the following:
•
The digital filter is a moving average (boxcar) filter. The filter rate sets
the number of readings that will be averaged. A reading is taken and
included in the average with each trigger. You must, therefore, supply
enough triggers to fill up the average before the filter will settle. If
your measurement requires averaging readings, use the
math
operations (Stats) instead of the digital filter (
•
The digital filter average may reset, based upon the measurement
value (
see page
57). You must use the
STATus:OPERation:CONDition
register to determine if the filter is settled with each reading taken.
•
The analog filter significantly slows the reading rate.
The input filters are disabled following a remote interface reset (
*RST
) or
preset (
SYStem:PREset
).
Chapter 4 Remote Interface Reference
Input Filter Commands
146
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......