Null (Relative) Commands
The meter uses independant null values for channel 1 and channel 2
voltage measurements, resistance measurements, and temperature
measurements.
See page 77 in Chapter 3
.
[SENSe:] | SENSe1: | SENSe2
VOLTage[:DC]:NULL[:STATe] {OFF | ON}
VOLTage[:DC]:NULL[:STATe]?
VOLTage[:DC]:NULL:VALue {<
value> | MIN | MAX}
VOLTage[:DC]:NULL:VALue?
These commands enable or disable the null state for voltage
measurements on the selected input channel. Null must be enabled
before the null value can be set. The null value set must be
within
±
132 V on each input channel.
[SENSe:]
FRESistance | RESistance
:NULL[:STATe] {OFF | ON}
:NULL[:STATe]?
:NULL:VALue {<
value> | MIN | MAX}
:NULL:VALue?
These commands enable or disable the null state for resistance
measurements. Null must be enabled before the null value can be set. The
null value set must be within
±
1.2 M
Ω
. Null is common to both 2-wire and
4-wire resistance measurements, that is, enabling null on 4-wire ohms
measurements also enables null on 2-wire ohms measurements.
[SENSe:]
TEMPerature
:NULL[:STATe] {OFF | ON}
:NULL[:STATe]?
:NULL:VALue {<
value> | MIN | MAX}
:NULL:VALue?
These commands enable or disable the null state for temperature
measurements. Null must be enabled before the null value can be set.
The null value set must be in the range of -300
°
C to 2000
°
C. Null is
common to all temperature measurements, that is, enabling null for one
probe type enables null for all probe types.
[SENSe:]NULL {OFF | ON | ONCE}
This command enables or disables null for the active channel or function.
OFF or ON disables or enables null using a value set in the null register.
ONCE enables null and makes the next reading the null value.
4
Chapter 4 Remote Interface Reference
Null (Relative) Commands
145
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......