Simplified Programming Sequence
You can program the meter to take measurements from the remote
interface using the following seven-step sequence.
Throughout this manual, the following conventions are used for SCPI
command syntax. Square brackets (
[]
) indicate optional keywords or
parameters. Braces (
{}
) enclose a list of parameters separated with a
vertical bar ( | ), use only one of the values . Triangle brackets (
<>
)
indicate that you must substitute a value for the enclosed parameter.
1. Place the meter in a known state (often the
reset
state).
2. Change the meter’s settings to achieve the desired configuration.
3. Set-up the triggering conditions.
4. Initiate or arm the meter for a measurement.
5. Trigger the meter to make a measurement.
6. Retrieve the readings from the output buffer or internal memory.
7. Read the measured data into your bus controller.
The
MEASure?
and
CONFigure
commands provide the most straight-
forward method to program the meter for measurements. You can select
the measurement function, range, and resolution all in one command.
The meter uses defaults for other measurement parameters.
Integration time, range, and resolution are local to the function or
channel used in the
MEASure?
and
CONFigure
commands. For example,
if you use the MEASure command to make a resistance measurement,
the range, resolution, and integration time set for voltage measurements
are not affected.
Other measurement parameters affect all measurement functions. When
you use the
MEASure?
and
CONFigure
commands the meter uses default
parameters. The default measurement parameters, common to all
functions or channels, used by the meter are shown in the table on the
next page.
Using the digital filter is
not
recommended from the remote interface,
➡
Chapter 4 Remote Interface Reference
Simplified Programming Sequence
126
Summary of Contents for 34420A
Page 1: ......
Page 2: ......
Page 13: ...1 Quick Start ...
Page 31: ...2 Front Panel Operation ...
Page 55: ...3 Features and Functions ...
Page 117: ...4 Remote Interface Reference ...
Page 200: ...5 Error Messages ...
Page 213: ...6 Application Programs ...
Page 245: ...6 Chapter 6 Application Programs Microsoft Excel Macro Example 245 ...
Page 246: ...Chapter 6 Application Programs Microsoft Excel Macro Example 246 ...
Page 247: ...7 Measurement Tutorial ...
Page 274: ...8 Specifications ...
Page 279: ...Dimensions Chapter 8 Specifications 34420A Nano Volt Micro Ohm Meter 280 ...
Page 294: ......