Installing Cryogenic Oven Cooling Accessory 19239A/B
Prepare the GC
2
5.
Remove the injection port cover by grasping its back edge and lifting it
upward. If an autosampler is installed, the injection port cover will not
be present.
6.
If an autosampler is installed on the instrument, it will be necessary to
remove it and its mounting bracket to allow removal of the left side
cover.
a.
Remove the autosampler tray from its mounting bracket by
simultaneously lifting and turning the two tray locks that hold it in
position, then sliding the tray away from the instrument.
b.
Lift the autosampler tray from its mounting bracket and set it aside.
c.
Remove the autosampler bracket by removing the six screws
securing it to the instrument.
Injection port cover
Top cover
Lid shaft
1/4-inch screw
and washer
Press here
to free cover