Installing Cryogenic Oven Cooling
Accessory 19239A/B

This manual describes how to install the following cryogenic oven cooling 

accessories:

19239A – Cryogenic Oven Cooling, CO

2

19239B – Cryogenic Oven Cooling, N

2

Prepare the GC

WARNING

Hazardous voltages are present in the instrument whenever the power cord 
is connected. Avoid a potentially dangerous shock hazard by disconnecting 
the power cord before working on the instrument.

1.

Set the main power line switch to the off position.

2.

Disconnect the power cable from its receptacle.

3.

Allow time for the oven and heated zones to cool.

4.

When the heated zones are cool, turn off all gas supplies at the source.

Red O visible

OFF

ON

Summary of Contents for 19239A

Page 1: ...Installation Guide Cryogenic Oven Cooling Accessory 19239A B...

Page 2: ...unit against unintended operation Refer servicing to qualified service personnel Substituting parts or performing any unauthorized modification to the instrument may result in a safety hazard Disconn...

Page 3: ...ven Cooling N2 Prepare the GC WARNING Hazardous voltages are present in the instrument whenever the power cord is connected Avoid a potentially dangerous shock hazard by disconnecting the power cord b...

Page 4: ...e it and its mounting bracket to allow removal of the left side cover a Remove the autosampler tray from its mounting bracket by simultaneously lifting and turning the two tray locks that hold it in p...

Page 5: ...two screws securing the left side panel along its bottom edge 8 Slide the left side panel towards the rear of the instrument and lift 9 Remove the right side panel by removing four screws two each al...

Page 6: ...stalling Cryogenic Oven Cooling Accessory 19239A B Prepare the GC 4 10 Remove the back cover of the instrument by removing four screws and sliding the cover off the rear of the instrument Screws Screw...

Page 7: ...bag 1 Using a screwdriver pierce the oven insulation covering the cutout for the cryo nozzle 2 Cover the open end of the cryo nozzle tubing then carefully insert the cryo valve assembly so that the no...

Page 8: ...ooling Accessory 19239A B Install the Cryogenic Valve 6 Insert lead pins into the plug connected at receptacle J9 disconnect the plug then insert the pins in the cryo locations on J9 J9 Location 4890...

Page 9: ...Installing Cryogenic Oven Cooling Accessory 19239A B Install the Cryogenic Valve 7 4 Position the pieces of insulation into the cryogenic valve cavity as shown J9 Location 5890 GC Red J9 Heaters...

Page 10: ...NG The plastic coated side tags faces toward the valve Restore Your GC to Operating Condition 1 Reinstall the covers 2 Connect the cryo valve to your cryogenic source as described in your Getting Star...

Page 11: ......