Replacing Assemblies
To remove and replace the front panel assembly
1
Perform previous procedures:
•
“To prepare the instrument for disassembly"
•
“To remove and replace the cover"
2
Disconnect the following cables from the module interface
board:
•
Front panel cable (16902/03A only)
•
Power On|Off cable
•
CD- ROM cable
3
Using a Torx T10 screwdriver, remove 4 screws that
secure the front panel to the chassis.
4
Remove/replace the front panel assembly.
5
Reverse this procedure to install the front panel assembly.