![Agilent Technologies 16900 Series Service Manual Download Page 49](http://html1.mh-extra.com/html/agilent-technologies/16900-series/16900-series_service-manual_2867838049.webp)
49
Chapter 3: Testing Logic Analyzer Performance
Test Pod 1 in 250 Mb/s Mode
9
Select
OK
to close the New Marker window.
Determine PASS/FAIL (2 of 2 tests)
Pass/Fail Point: The Listing window is set up to search for the appropriate
number of A's and 5's in the acquisition. If the logic analyzer does not
detect the correct number of A’s and 5’s, an error window will appear.
1
Select the Run Repetitive icon
. Let the logic analyzer run for about one
minute. The analyzer will acquire data and the Listing Window will
continuously update.
If the “can’t find occurrence” window appears, then the logic analyzer fails
the test.
Check your test setup. If the failure is not the result of a problem with the
test setup, record the failure in the “
Test 2 of 2: Correct number of
occurrences detected
” section of the Performance Test Record (page 60).
NOTE:
Be sure that the black ground lead is making good contact with the ground
pin on the test connector.
2
When about one minute has elapsed, select the
Stop
button
to stop the
acquisition.
If the “can’t find occurrence” window does not appear, then the analyzer
passes the test. Record “Pass” in the “
Test 2 of 2: Correct number of
occurrences detected
” section of the Performance Test Record (page 60).
Summary of Contents for 16900 Series
Page 3: ...3 Chapter The 16910A Logic Analyzer The 16911A Logic Analyzer...
Page 8: ...8 Contents...
Page 14: ...14 Chapter 1 General Information...
Page 18: ...18 Chapter 2 Preparing for Use...
Page 61: ...61 4 Calibrating This chapter gives you instructions for calibrating the logic analyzer...
Page 65: ...65 Chapter 5 Troubleshooting Troubleshooting Flowchart 1...
Page 66: ...66 Chapter 5 Troubleshooting Troubleshooting Flowchart 2...
Page 82: ...82 Chapter 5 Troubleshooting To test the cables 18 Return to the troubleshooting flow chart...
Page 94: ...94 Chapter 7 Replaceable Parts 16910A Exploded View Exploded view of the 16910A logic analyzer...
Page 95: ...95 Chapter 7 Replaceable Parts 16911A Exploded View Exploded view of the 16911A logic analyzer...
Page 96: ...96 Chapter 7 Replaceable Parts...
Page 102: ...102 Index...