
Agilent 1290 Infinity LC System Manual and Quick Reference
105
Quick Start Guide
5
Data Analysis
Integrating a Signal
1
Select the Integration Task Tool (see figure below). The
Integrate
icon and
the
Set Integration Events Table
icon are highlighted in the figure shown below.
2
Click the
Set Integration Events Table
icon to open the table as shown.
3
Set
Baseline Correction
to
Advanced
for gradient runs.
4
Set
Slope Sensitivity
to 50. Higher numbers will integrate steeper peaks and
ignore less steep peaks.
5
Set the
Peak Width
value to the narrowest peak of interest, approximately
0.01 in this case.
6 Area Reject
and
Height Reject
can be set to reject the smallest peaks.
7
Click the
Integrate
icon to update the results using these new settings.