Compatibility Matrix
Supported Chromatographic Data Systems
2D-LC User Guide
46
Supported Chromatographic Data Systems
Following revision of MassHunter Workstation Data Acquisition is
recommended:
•
MassHunter Workstation Data Acquisition 11 for Q-TOF/TOF (or higher)
MassHunter Workstation Data Acquisition and Q-TOF/TOF instruments can be
controlled with Agilent driver-based 2D-LC Solution. Please see the
CDS_requirements in the CDS document folder which LC modules are supported.
This software has been tested successfully with 12 LC modules. Please note that
complex systems can increase memory consumption in MassHunter, which may
decrease system stability. This is very unlikely, but it is still advisable to consider
the following:
•
Restart MassHunter Workstation Software from time to time, e.g. once per
week or more often for complex systems
•
Perform data analysis, reporting, online help reading in an offline copy of the
MassHunter instrument
•
Save data before starting new tasks
•
Avoid high levels of interactivity during runs by editing methods, changing
signal plots settings, etc.