12
Maintenance
Replacing the Fuses of the Infinity Valve Drive
2D-LC User Guide
332
Replacing the Fuses of the Infinity Valve Drive
When
If the flow module shows no reaction.
Tools required
Description
Screwdriver
Parts required
#
p/n
Description
2
2110-1486
Fuse 2 AT250 V
WARNING
Electrical shock
Disconnect the module from line power before changing a fuse or trying
to open the hatch of the power input socket.
Never reconnect the line power before havingthe power input socket
closed.
1
Switch off the instrument. The line switch is located at
the rear of the module.
2
Disconnect the power cable from the power input
socket.