![Agilent Technologies 1260 Infinity II User Manual Download Page 153](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii/1260-infinity-ii_user-manual_2867537153.webp)
Agilent 1260 Infinity II Preparative Autosampler User Manual
153
Maintenance
9
Remove and Install Doors
Remove and Install Doors
When
If the front door is defective or the hinge are damaged.
Tools required
Description
Flat screwdriver
Parts required
p/n
Description
G7129-68702
Door Assy
Preparations
Finish any pending acquisition job and return any plate on the workspace back to the hotel.
1
Press the release buttons and pull the front door out.
2
For the Installation of the front door, insert the hinges
into their guides and move the door in until the release