Agilent 1260 Infinity II Preparative Autosampler User Manual
151
Maintenance
9
Overview of Maintenance
Overview of Maintenance
It is necessary to perform periodic inspection of this instrument to ensure its
safe use. It is possible to have these periodic inspections performed by Agilent
service representatives on a contractual basis. For information regarding the
maintenance inspection contract, contact your Agilent representative.
The following pages describe the maintenance (simple repairs) of the module
that can be carried out without opening the main cover.
Table 11
Overview of maintenance
Procedure
Typical interval (minimum)
Notes
Change needle/needle seat
30000 needle into seat movements
Change peristaltic pump
cartridge
3000 min on time
Change rotor seal
30000 injections