Agilent 1260 Infinity II Preparative Autosampler User Manual
113
Troubleshooting and Diagnostics
7
Maintenance Functions
Change Metering Seal or Piston
The change-metering seal/piston function draws the piston away from the
home position, relieving the tension on the spring. In this position, the
analytical head assembly can be removed and reinstalled easily after
maintenance.
User Interface
The commands for the control software are:
Start
Draws the piston away from the home position, relieving the tension on the
spring.
Next
Repositions the piston at the home position.
Back
Completes the procedure.
Using the
Change Seal
Function
1
Select
Start
to move the piston to the maintenance position.
2
Exchange the metering seal (see
“Exchange the Metering Seal and
3
Select
Next
to move the piston back to the home position.
4
Select
Back
to finish the procedure.