Agilent InfinityLab LC Series Vialsamplers User Manual
91
Installation of the Optional Sample Cooler or Thermostat
4
Install the Sample Cooler or Thermostat
Install the Sample Cooler or Thermostat
Parts required
#
p/n
Description
1
Sampler
1
G7167-60005
Sample cooler
OR
1
G7167-60101
InfinityLab Sample Thermostat
1
Power cord
1
5067-6208
Condensate Drainage Kit
1
G7167-90171
Installation of the Infinity II Cooler/Thermostat Condensate
Drainage Tubing Kit
Preparations
Sampler is installed in the stack.
WA R N I N G
Flammable refrigerant
Formation of flammable gas-air mixtures inside the sample thermostat and
laboratory.
➔
Keep open fire or sources of ignition away from the device.
➔
Ensure a room size of 4 m
3
(1 m
3
for every 8 g of R600a refrigerant inside of the
sample thermostat).
➔
Ensure adequate ventilation: typical air exchange of 25 m
3
/h per m
2
of laboratory
floor area.
➔
Keep all ventilation openings in the enclosure clear of obstructions. Do not block
the openings on the circumference of the sample thermostat.
WA R N I N G
Flammable refrigerant used
➔
When handling, installing and operating the Sample Thermostat, care should be
taken to avoid damage to the refrigerant tubing or any part of the Sample
Thermostat.