Agilent InfinityLab LC Series Vialsamplers User Manual
3
Using the Module
Remove a Column (ICC)
Remove a Column (ICC)
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Parts required
p/n
Description
Column (up to 30 cm length)
5067-5957
InfinityLab Quick Connect Assy ST 0.12 mm x 105 mm
5067-5966
InfinityLab Quick Turn fitting
Preparations
•
Switch off the pump
•
Switch off the sampler
•
Open the doors of the sampler
1
Push the latch.
2
Open the flap 90 °.
3
Remove the column.
4
Unscrew the right fitting.