![Agilent Technologies 1260 Infinity II Series User Manual Download Page 60](http://html.mh-extra.com/html/agilent-technologies/1260-infinity-ii-series/1260-infinity-ii-series_user-manual_2867552060.webp)
60
Agilent InfinityLab LC Series Vialsamplers User Manual
3
Using the Module
Install the Integrated Column Compartment
Install the Integrated Column Compartment
Tools required
p/n
Description
8710-0510
Open-end wrench 1/4 — 5/16 inch
Parts required
#
p/n
Description
1
G7130-60030
ICC Column Heater 3 µL volume
OR
1
G7130-60060
ICC Column Heater 6 µL volume
Preparations
•
Switch off the pump
•
Switch off the sampler
•
Column is installed in the ICC
•
Open the doors of the sampler
WA R N I N G
Fire or explosion due to missing leak drainage
Solvents may run into the module if Column Shelf or Integrated Column
Compartment are not installed.
➔
Install either Column Shelf or Integrated Column Compartment before starting the
module.