Site Requirements and Specifications
2
Specifications of the Sample Cooler
Specifications of the Sample Cooler
Cooling unit is designed as vapor-compression refrigeration system. Contains
fluorinated greenhouse gas (refrigerant) according to the Kyoto protocol. For
specifications of refrigerant, charge capacity, carbon dioxide equivalent
(CDE), and global warming potential (GWP) see instrument label.
Table 5
Physical Specification of the Sample Cooler
Type
Specification
Comments
Weight <
6 kg
Dimensions
(height × width × depth)
205 mm x 340 mm x 370 mm
Refrigerant gas
HFC-134a (0.042 kg)
Ozone depletion
potential (ODP) = 0
Supply voltage
24 VDC (nominal)
Current
10 A max.
Ambient operating temperature
4 – 40 ° C (39.2 – 104 ° F)
Ambient non-operating
temperature
-40 – 70 ° C (-20 – 158 ° F)
Humidity
< 95 % r.h. at 40 °C (104 °F)
Non-condensing
Operating altitude
Up to 3000 m (9842 ft)
Non-operating altitude
Up to 4600 m (15091 ft)
Safety standards:
IEC, EN, CSA, UL
Installation category II, Pollution
degree 2
For indoor use only.
ISM Classification
ISM Group 1 Class B
According to CISPR 11