Agilent InfinityLab LC Series Vialsamplers User Manual
41
Site Requirements and Specifications
2
Performance Specifications
Performance Specifications
Performance Specifications (G7129A)
Table 2
Performance Specifications 1260 Infinity II Vialsampler (G7129A)
Type
Specification
Comment
Injection range
0.1 – 100 µL in 0.1 µL increments with 100 µL
up to 60 MPa
0.1 – 900 µL in 0.1 µL increments with 900 µL
up to 40 MPa
Up to 1800 µL with multiple draw (hardware
modification required)
Requires 900 µL analytical head
Precision
<
0.25 % RSD of peak areas from 5 µL to 100 µL
Measured caffeine
Pressure range
0 – 60 MPa (0 – 600 bar, 0 – 8702 psi)
0 – 40 MPa (0 – 400 bar, 0 – 5801 psi)
for 900 µL Analytical Head
Sample viscosity
range
0.2 – 5 cp
Sample capacity
132 x 2 mL vial (two trays default)
100 x 2 mL vial (two classic trays optional)
36 x 6 mL vials (two trays optional)
Carry over
<
0.004 % (40 ppm) with needle wash
Injection cycle time
18 s for draw speed 200 µL/min
Ejection speed: 200 µL/min
Injection volume: 1 µL
Minimum sample
volume
1 µL from 5 µL sample in 100 µL microvial, or
1 µL from 10 µL sample in 300 µL microvial.
Needle height offset has to be adapted to
ensure that needle doesn’t touch vial bottom.
Default needle height = 0 equates to 2 mm
above the vial bottom.
Instrument Control
LC & CE Drivers A.02.12 or above
Instrument Control Framework (ICF) A.02.03 or
above
Instant Pilot (G4208A) with firmware B.02.19 or
above
Lab Advisor B.02.07 or above
For details about supported software versions
refer to the compatibility matrix of your version
of the LC and CE Drivers