Agilent InfinityLab LC Series Vialsamplers User Manual
8
Error Information
Integrated Column Compartment (ICC) Heater Error Messages
ICC Heater Temp Sensor Failure
Error ID: 30821, 30823,30825
The Integrated Column Compartment (ICC) Heater has detected a reading
issue with the heater sensors in the heating element.
ICC Heater Over Temp Failure
Error ID: 30824
The Integrated Column Compartment (ICC) Heater has detected a
temperature reading above 100 °C. This exceeds the limit of the heating
element.
Probable cause
Suggested actions
1
Defective cable or connector
Please contact your Agilent service
representative.
2
Defective ICC heater sensor
Replace the ICC heater.
Probable cause
Suggested actions
1
Defective cable or connector
Please contact your Agilent service
representative.
2
Defective ICC heater
Replace the ICC heater.