Agilent InfinityLab LC Series Vialsamplers User Manual
179
Error Information
8
Sampler Error Messages
Valve switch not found.
Error ID: 34319
The injection valve has failed to switch into the position. The valve position
sensor monitors the position of the valve. If the valve fails to switch into the
position (mainpass (main path) or bypass), or if the sensor fails to recognize
the valve position, the error message is generated.
Metering Initialization Failed
Error ID: 34322
The metereing device has failed to do the initialization. If the piston fails to
move to the home position, or if the metering device is not recognized by the
instrument.
Probable cause
Suggested actions
1
Defective rotor seal/stator head.
Exchange the rotor seal/stator head.
2
Defective injection valve.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Metering device is not recognized.
Configure the metering device in LabAdvisor
2
Loose belt.
Please contact your Agilent service
representative.
3
Rear wheel in the hydraulic unit is loose.
Please contact your Agilent service
representative.
4
Defective metering-drive motor.
Please contact your Agilent service
representative.
5
Defective metering device
Please contact your Agilent service
representative.