Agilent InfinityLab LC Series Vialsamplers User Manual
8
Error Information
Sampler Error Messages
Sampler Error Messages
Defective Gripper Assembly
Valve Failed
Error ID: 34219, 34220
The injection valve failed to switch to the mainpass (main path) position.
The switching of the injection valve is monitored by two microswitches on the
valve assembly. The switches detect the successful completion of the valve
movement. If the valve fails to reach the mainpass (main path) position, or if
the microswitch does not close, the error message is generated.
Probable cause
Suggested actions
1
Mechanical obstruction.
Ensure unobstructed movement of the transport
assembly.
2
High friction in the transport assembly.
Please contact your Agilent service
representative.
3
Defective motor assembly.
Please contact your Agilent service
representative.
4
Defective sample transport assembly flex
board.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.
Probable cause
Suggested actions
1
Defective injection valve.
Please contact your Agilent service
representative.
2
Defective main board.
Please contact your Agilent service
representative.