Agilent InfinityLab LC Series Vialsamplers User Manual
7
Troubleshooting and Diagnostics
Step Commands
Needle Up
Lifts the needle out of the vial.
Command also switches the
valve to bypass if it is not
already in that position.
Vial to Tray
Returns the selected vial to the tray
position.
Command also lifts the
needle to the upper position.
Needle into Seat
Lowers the needle arm into the seat.
Command also returns the
vial to the tray position.
Valve Mainpass
Switches the injection valve to the
mainpass (main path) position.
Reset
Resets the injector.
Table 15
Injector step commands
Step
Action
Comments