Theory of Operation
8
Optional Interface Boards
The Agilent 1200 Series modules have one optional board slot that allows
addition of an interface board to the modules.
BCD Board
The BCD board provides a BCD output for the bottle number of the Agilent
1200 Series autosampler and four external contacts. The external contact
closure contacts are relay contacts. The maximum settings are: 30 V (AC/DC);
250 mA (fused).
Table 43
Optional Interface Boards
Description
Part Number
BCD Board
G1351-68701
Fuse 250 mA (four are on the board)
2110-0004
LAN Communication Interface Board
G1369A or G1369-60001
Figure 41
Block Diagram BCD Board
Processor
interface
Board
identification
BCD register
External
contacts
Line driver
12
4 x
250 mA
BCD
connector
External contact
connector
RFI
fi
lt
er
RFI
fil
ter
+