Replacing Major Assemblies
8
11713B/C Operating and Service Manual
103
Power Supply Module Removal/Replacement
To Remove Power Supply Module
1
Remove top clamshell (M2), refer
2
Remove screws (F1), spring washer (F2) and flat washer (F3) to
release PSU cover (F4)
3
Remove screws (F5) that secure power supply module (F6) to top
clamshell (M2)
4
Now you can remove power supply module (F6)
To Replace Power Supply Module
1
Place power supply module on top clamshell (M2), ensure you
position the power supply where the 8- pin connector face the
opposite of PSU desk’s lip
2
Secure power supply module (F6) with screws (F5)
3
Secure PSU cover (F4) with screws (F3), spring washer (F2) and
flat washer (F1)
4
Replace top clamshell (M2), refer to
C A U T I O N
This module contains components that can be damaged or destroyed by
electrostatic discharge. It should be serviced at a static-safe workstation.
Refer to “preparing a static-safe workstation” in chapter 9.
Power supply module (F6)
PSU cover (F4)
Top clamshell (M2)
Summary of Contents for 11713B
Page 1: ...Agilent Technologies Agilent 11713B C Attenuator Switch Drivers Operating and Service Manual ...
Page 12: ...12 11713B C Operating and Service Manual ...
Page 24: ...24 11713B C Operating and Service Manual 1 Introduction ...
Page 43: ...Specifications 3 11713B C Operating and Service Manual 43 Declaration of Conformity ...
Page 44: ...44 11713B C Operating and Service Manual 3 Specifications ...
Page 64: ...64 11713B C Operating and Service Manual 5 Local Operations ...
Page 88: ...88 11713B C Operating and Service Manual 7 Remote Interface Configurations ...
Page 104: ...104 11713B C Operating and Service Manual 8 Replacing Major Assemblies ...
Page 114: ...114 11713B C Operating and Service Manual 9 Servicing the Attenuator Switch Driver ...