The following symptoms may indicate a problem with the
probe or the way it is used. The probe is a high frequency
device with many critical relationships between parts. For
example, the frequency response of the amplifier on the
hybrid is trimmed to match the output coaxial cable. As a
result, to return the probe to optimum performance requires
factory repair. If the probe is under warranty, normal
warranty services apply.
Probe Calibration Fails
Probe calibration failure with an oscilloscope is usually
caused by improper setup. If the calibration will not pass,
check the following:
■
Check that the probe passes a waveform with the correct
amplitude.
■
If the probe is powered by the oscilloscope, check that the
offset is approximately correct. The probe calibration
cannot correct major failures.
■
Be sure the oscilloscope passes calibration without the
probe.
■
Be sure that the probe head that you are using has been
in the oscilloscope’s Probe Setup dialog box.
Incorrect Pulse Response
(flatness)
If the probe's pulse response shows a top that is not flat,
check for the following:
■
Output of probe must be terminated into a proper 50
Ω
termination. If you are using the probe with an Infiniium
oscilloscope, this should not be a problem. If you are
using the probe with other test gear, ensure the probe is
terminated into a low reflectivity 50
Ω
load (~ ±2%).
Summary of Contents for 1168/9A
Page 1: ...Agilent Technologies Agilent 1168 9A Differential and Single Ended Probes User s Guide ...
Page 6: ...6 1168 9A Probes User s Guide Contents ...
Page 9: ...Getting Started 1 Introduction 1168 9A Probe s User s Guide 9 Figure 2 Available Probe Heads ...
Page 40: ...40 1168 9A Probe s User s Guide 1 Getting Started Service ...
Page 104: ...104 1168 9A Probes User s Guide 2 Using Probe Heads N2887A 8A Soft Touch Probe Heads ...
Page 120: ...120 1168 9A Probes User s Guide 3 Maintaining Probe Heads E2677A 9A Solder In Probe Heads ...
Page 150: ...150 1168A 9A Probe s User s Guide 5 Characteristics and Specifications Probe Dimensions ...
Page 202: ...202 1168A 9A Probe s User s Guide 6 Performance Plots N2887A N2888A Soft Touch Probe Heads ...
Page 250: ...250 1168A 9A Probe s User s Guide 9 Replacement Parts Other Accessories ...