Calibrating the Probe
Calibration Test Record
Calibration Test Record
1141A/1142A Differential Probe
Tested by_________________________
Serial No. ______________________________
Work Order No.____________________
Recommended Test Interval - 1 Year/2000 hours
Date____________________
Recommended next testing_________________
Temperature_____________
Test
Limits
Results
dc Gain Accuracy
Probe Only
+0.98 mV to +1.02 mV
10x
+0.096 mV to +0.104 mV
100x
+0.0096 mV to 0.0104 mV
Bandwidth
> 0.707 at 200 MHz
CMRR
1 MHz
≥
3000:1
100 MHz
≥
10:1