1100 Series Nano Pump Service Manual
83
Troubleshooting and Test Functions
2
Fourth plateau negative or unstable and at least one other plateau positive
Potential Cause
Corrective Action
Air in pump chamber of channel B or seals not
yet seated.
Flush channel B thoroughly with isopropanol
under pressure (restriction capillary).
Loose active inlet valve in channel B.
Tighten the active inlet valve in channel B
(14mm wrench). Do not overtighten!
Loose pump head screws in channel B.
Ensure the pump head screws in channel B are
tight.
Loose outlet valve in channel B.
Ensure the sieve in the outlet valve is installed
correctly. Tighten the outlet valve.
Leaking seal or scratched plunger in channel B1.
Exchange the pump seals in channel B. Check
the plungers for scratches. Exchange if
scratched.
Defective active inlet valve in channel B.
Exchange the active inlet valve in channel B.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......