78
1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
No pressure increase or minimum pressure of plateau 1 not reached
Potential Cause
Corrective Action
Pump not running.
Check the logbook for error messages.
Wrong solvent-line connections to solvent
selection valve.
Ensure the solvent lines from the degasser to
the solvent selection valve are connected
correctly.
Loose or leaky fittings.
Ensure all fittings are tight, or exchange
capillary.
Large leaks (visible) at the pump seals.
Exchange the pump seals.
Large leaks (visible) at active inlet valve, outlet
valve.
Ensure the leaky components are installed
tightly. Exchange the component if required.
Pressure limit not reached but plateaus horizontal or positive
Potential Cause
Corrective Action
Degasser and pump channels A and/or B not
flushed sufficiently (air in the channels).
Purge the degasser and pump channels
thoroughly with isopropanol under pressure
(use the restriction capillary).
Wrong solvent.
Install isopropanol. Purge the degasser and
pump channels thoroughly.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......