1100 Series Nano Pump Service Manual
77
Troubleshooting and Test Functions
2
Evaluating the Results of the Leak Test
Defective or leaky components in the pump head lead to changes in the
leak-test pressure plot. Typical failure modes are described below.
Figure 10
Leak Test Pressure Plot
Plateau 1 Plateau 2
Plateau 3
Plateau 4
Time [minutes]
Pressure
[bar]
N O T E
Please notice the difference between an error in the test and a failure of the test! An error
means that during the operation of the test there was an abnormal termination. If a test
failed, this means that the results of the test where not within the specified limits.
N O T E
Often it is only a damaged blank nut itself (poorly shaped from overtightening) that causes
a failure of the test. Before investigating on any other possible sources of failure make sure
that the blank nut you are using is in good condition and properly tightened!
N O T E
The pressure plots shown below are examples only. The plots may vary depending on the
type and degree of leakage.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......