1100 Series Nano Pump Service Manual
57
Troubleshooting and Test Functions
2
Servo Restart Failed
Servo Restart Failed:
Pump channel A
B: Servo Restart Failed:
Pump channel B
The pump motor in the nano pump was unable to move into the correct
position for restarting.
When the nano pump is switched on, the first step is to switch on the C phase
of the variable reluctance motor. The rotor should move to one of the C
positions. The C position is required for the servo to be able to take control of
the phase sequencing with the commutator. If the rotor is unable to move, or if
the C position cannot be reached, the error message is generated.
Probable Causes
•
Disconnected or defective cables.
•
Mechanical blockage of the nano pump.
•
Defective pump drive assembly.
•
Defective CSM board.
Suggested Actions
✔
Ensure the pump assembly cables are not damaged or dirty. Make sure the
cables are connected securely to the CSM board.
✔
Remove the pump-head assembly. Ensure there is no mechanical blockage
of the pump-head assembly or pump drive assembly.
✔
Exchange the pump drive assembly.
✔
Exchange the CSM board.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......