2
Electro-Magnetic-Proportional-Valve (EMPV) Missing
EMPV Missing
The EMPV in the micro pump is missing or defective.
Probable Causes
•
Disconnected or defective cable.
•
Defective solenoid.
Suggested Actions
✔
Ensure the connection cable is seated correctly.
✔
Exchange the solenoid of the EMPV.