Troubleshooting and Test Functions
Inlet-Valve Missing
Inlet-Valve Missing:
Pump channel A
B: Inlet-Valve Missing:
Pump channel B
The active-inlet valve in the nano pump is missing or defective.
The processor checks the presence of the active-inlet valve connector every
2 seconds. If the connector is not detected by the processor, the error message
is generated.
Probable Causes
•
Disconnected or defective cable.
•
Disconnected or defective connection cable (front panel to CSPM board).
•
Defective active inlet valve.
Suggested Actions
✔
Ensure the pins of the active inlet valve connector are not damaged. Ensure
the connector is seated securely.
✔
Ensure the connection cable is seated correctly. Exchange the cable if
defective.
✔
Exchange the active inlet valve.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......