Temperature Out of Range
Temperature Out of Range 0:
Pump channel A
Temperature Out of Range 1:
Pump channel B
One of the temperature sensor readings in the motor-drive circuit are out of
range.
The values supplied to the ADC by the hybrid sensors must be between 0.5 V
and 4.3 V. If the values are outside this range, the error message is generated.
Probable Causes
•
Defective CSM board.
Suggested Actions
✔
Exchange the CSM board.