1100 Series Nano Pump Service Manual
47
Troubleshooting and Test Functions
2
Inlet-Valve Fuse
Inlet-Valve Fuse 0:
Pump channel A
Inlet-Valve Fuse 1:
Pump channel B
One of the active-inlet valves in the nano pump has drawn excessive current
causing the inlet-valve electronic fuse to open.
Probable Causes
•
Defective active inlet valve.
•
Defective connection cable (front panel to CSM board).
•
Defective CSM board.
Suggested Actions
✔
Restart the nano pump. If the error message appears again, exchange the
active inlet valve.
✔
Exchange the connection cable.
✔
Exchange the CSM board.