46
1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Valve Fuse
Valve Fuse 0:
Channels A1 and A2
Valve Fuse 1:
Channels B1 and B2
One of the solvent-selection valves in the nano pump has drawn excessive
current causing the selection-valve electronic fuse to open.
Probable Causes
•
Defective solvent selection valve.
•
Defective connection cable (front panel to CSM board).
•
Defective CSM board.
Suggested Actions
✔
Restart the nano pump. If the error message appears again, exchange the
solvent selection valve.
✔
Exchange the connection cable.
✔
Exchange the CSM board.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......