1100 Series Nano Pump Service Manual
31
Troubleshooting and Test Functions
2
Leak
A leak was detected in the nano pump.
The signals from the two temperature sensors (leak sensor and
board-mounted temperature-compensation sensor) are used by the leak
algorithm to determine whether a leak is present. When a leak occurs, the leak
sensor is cooled by the solvent. This changes the resistance of the leak sensor
which is sensed by the leak-sensor circuit on the CSM board.
Probable Causes
•
Loose fittings.
•
Broken capillary.
•
Loose or leaking active inlet valve, outlet ball valve, or EMPV.
•
Defective pump seals.
Suggested Actions
✔
Ensure all fittings are tight.
✔
Exchange defective capillaries.
✔
Ensure pump components are seated correctly. If there are still signs of a
leak, exchange the appropriate seal (active inlet valve, outlet ball valve).
✔
Exchange the pump seals.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......