28
1100 Series Nano Pump Service Manual
2
Troubleshooting and Test Functions
Shutdown
An external instrument has generated a shut-down signal on the remote line.
The nano pump continually monitors the remote input connectors for status
signals. A LOW signal input on pin 4 of the remote connector generates the
error message.
Probable Causes
•
Leak detected in another module with a CAN connection to the system.
•
Leak detected in an external instrument with a remote connection to the
system.
•
Shut-down in an external instrument with a remote connection to the
system.
•
The degasser failed to generate sufficient vacuum for solvent degassing.
Suggested Actions
✔
Fix the leak in the external instrument before restarting the nano pump.
✔
Check external instruments for a shut-down condition.
✔
Check the vacuum degasser for an error condition. Refer to the
Reference
Manual
for the Agilent 1100 Series vacuum degasser.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......